7. Maintenance
Operator’s Manual
7-8
Cleaning
Caution
Wear appropriate personal protective clothing to prevent
infection when cleaning.
Clean the analyzer after use or periodically.
Do not spray cleaning solution directly onto the analyzer.
Do not allow cleaning solution to enter the analyzer.
Do not use force to wipe the screen.
Prepare 0.5% hypochlorite cleaning solution immediately before
use.
Dispose of all waste after cleaning in accordance with the
laboratory’s established procedures for disposing of biohazardous
materials.
Cleaning
Procedure
1.
Use 0.5% hypochlorite cleaning solution.
Note:
Commercial Clorox contains approximately 5% sodium
hypochlorite.
To prepare 0.5% hypochlorite solution, mix 1 part of Clorox
and 9 parts of water.
2.
Dampen a soft cloth with the cleaning solution.
3.
Using a dampened soft cloth, wipe sampler cover, screen, and
other contaminated areas on the analyzer.
4.
Allow to air-dry for about 10 minutes.
Using a soft cloth dampened with water, wipe the analyzer.
5.
Using a dry cloth, dry the surface of the analyzer.
Summary of Contents for i-Smart 300
Page 2: ...Operators should read the entire manual before installing and operating the analyzer...
Page 5: ...1 Introduction 1 Introduction Safety 1 1 Symbols and Terms 1 3 Product Description 1 5...
Page 12: ...1 Introduction 1 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 25: ...3 Analyzer Setting Operator s Manual 3 2 Screen Menus continued Menu options...
Page 48: ...3 Analyzer Setting 3 23 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 49: ...4 Sample Analysis 4 Sample Analysis Blood Samples 4 1 QC Samples 4 8...
Page 76: ...5 Database 5 13 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 77: ...6 Shutdown 6 Shutdown Analyzer Shutdown 6 1 Cartridge Restart 6 3...
Page 90: ...7 Maintenance 7 9 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 91: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 8 1 Error Code 8 6...
Page 98: ...8 Troubleshooting 8 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 104: ...9 Specification 9 5 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...