2. Installation
Operator’s Manual
2-4
Power-On,
continued
Turn Power
Switch On,
continued
5.
Check that the power cord icon
이
appears at the bottom
row of the screen.
Note:
If the power cord icon does not appear on the screen,
check for any loose power connections between the analyzer and
the electrical outlet.
If no loose power connection is found and the power cord icon
still does not appear, call a service engineer immediately.
6.
Check the date and time of the analyzer and, if necessary, change
date and/or time. Refer to
3. Analyzer Settings
Note:
Changing the date and time of the analyzer is only allowed
before cartridge is installed on the analyzer. Once the cartridge
is installed, the date and time cannot be changed until the
cartridge is removed.
Summary of Contents for i-Smart 300
Page 2: ...Operators should read the entire manual before installing and operating the analyzer...
Page 5: ...1 Introduction 1 Introduction Safety 1 1 Symbols and Terms 1 3 Product Description 1 5...
Page 12: ...1 Introduction 1 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 25: ...3 Analyzer Setting Operator s Manual 3 2 Screen Menus continued Menu options...
Page 48: ...3 Analyzer Setting 3 23 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 49: ...4 Sample Analysis 4 Sample Analysis Blood Samples 4 1 QC Samples 4 8...
Page 76: ...5 Database 5 13 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 77: ...6 Shutdown 6 Shutdown Analyzer Shutdown 6 1 Cartridge Restart 6 3...
Page 90: ...7 Maintenance 7 9 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 91: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 8 1 Error Code 8 6...
Page 98: ...8 Troubleshooting 8 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 104: ...9 Specification 9 5 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...