4. Sample Analysis
4-1
i-Smart 300 Blood Gas Analyzer
Blood Samples
Introduce
Sample
Note:
Sample analysis is available only when the analyzer is in
Ready
state. When the analyzer is not in
Ready
state, the sampler
cover cannot be lifted because it is locked.
1.
Check that the analyzer is in
Ready
state.
2.
Lift up the sampler cover. The following message will appear.
Note:
To cancel sampling, lower the sampler cover to the original
position before pressing
Aspirate
.
Once
Aspirate
is pressed, the sample analysis cannot be
cancelled.
Summary of Contents for i-Smart 300
Page 2: ...Operators should read the entire manual before installing and operating the analyzer...
Page 5: ...1 Introduction 1 Introduction Safety 1 1 Symbols and Terms 1 3 Product Description 1 5...
Page 12: ...1 Introduction 1 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 25: ...3 Analyzer Setting Operator s Manual 3 2 Screen Menus continued Menu options...
Page 48: ...3 Analyzer Setting 3 23 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 49: ...4 Sample Analysis 4 Sample Analysis Blood Samples 4 1 QC Samples 4 8...
Page 76: ...5 Database 5 13 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 77: ...6 Shutdown 6 Shutdown Analyzer Shutdown 6 1 Cartridge Restart 6 3...
Page 90: ...7 Maintenance 7 9 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 91: ...8 Troubleshooting 8 Troubleshooting Troubleshooting 8 1 Error Code 8 6...
Page 98: ...8 Troubleshooting 8 7 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...
Page 104: ...9 Specification 9 5 i Smart 300 Blood Gas Analyzer This page is intentionally left blank...