7. Maintenance
7-3
i-Smart 300 Blood Gas Analyzer
Calibration,
continued
Sensor State
1.
State of each sensor will appear as okay in blue background, drift
error in yellow background or slope error in red background at
the main screen based on the results of the last calibration.
2.
Press the
Sensor State
icon twice quickly. The last slope and
the allowable slope range of the sensor will appear.