Error Messages
If the power fails or any interruption occurs during a run, the LCD displays an error message.
Affected samples in the thermal block should not be used with the BAX
System. Remove and
dispose of these samples according to your standard operating procedures.
The letter “
F
” (Fault) appears in the lower
right corner of the LCD display if an
interruption occurs during any step in the
program.
This fault indicator appears on the LCD
display until the program completes or the
program is aborted by the user.
Affected samples should not be used with
the BAX System. Remove and dispose of
these samples according to your standard
operating procedures.
The
Power Failure
message appears when
power is interrupted or fails during a run.
Toggle the on/off switch on the back panel
to reset the unit.
Affected samples should not be used with
the BAX System. Remove and dispose of
these samples according to your standard
operating procedures.
Denaturing @95C
GmPos F
Gram Positive
POWER FAILURE!