20 

Automated Thermal Block 

13  Press the SELECT/CONTINUE push 

button until the arrow 

 is next to the 

Lo Temp Cal Pt

 option. 

Press the SCROLL and 

SELECT/CONTINUE buttons at the 

same time to select the 

Lo Temp Cal 

Pt

 option. 

14  The LCD display changes to a screen 

that compares the Displayed and 

Measured temperatures. 

If needed, use the SCROLL button to 

increase the value of the Measured 

temperature or the 

SELECT/CONTINUE button to 

decrease the value of the Measured 

temperature to the value displayed on 

the thermometer.

  

Note: When entering the Measured 

temperature, do not include decimal 

places.

 

15  Press the SCROLL and 

SELECT/CONTINUE buttons at the 

same time to set the Measured 

temperature.  

If necessary, the thermal block will 

automatically adjust the Plate 

Temperature to align with the Set Point.

  

16  Press the SCROLL button to increase 

the value of the set point to 95 °C. The 

block automatically begins the heating 

process. 

If needed, use the 

SELECT/CONTINUE button to 

decrease the value of the set point.

 

Note: Be sure to wait for the temperature 

reading of the thermometer to equilibrate 

before proceeding

17  Press the SCROLL and 

SELECT/CONTINUE buttons at the 

same time. The calibration menu again 

appears on the LCD display. 

18  Press the SELECT/CONTINUE button 

until the arrow 

 is next to the 

Calibrate

 option. 

Press the SCROLL and 

SELECT/CONTINUE buttons at the 

same time to select the 

Calibrate

 

option. 

Lo Temp Cal Pt

Hi Temp Cal Pt

 Displayed: 37C 

  Measured: 38C 

Set Point: 37C 

Plate Temp: 37C 

Set Point: 95C 

Plate Temp: 37C 

Exit

Calibrate

  Exit 

Calibrate

Summary of Contents for DUP-1000

Page 1: ...Model DUP 1000 User Guide...

Page 2: ...ts affiliates All other trademarks are the property of their holders FIELD OF USE The Automated Thermal Block is manufactured by Torrey Pines Scientific for use with the BAX System Please see BAX Syst...

Page 3: ...Front Panel Display 4 Program Menu 5 Turning the Thermal Block On and Off 5 Running the Gram Positive Program 6 Running the Gram Negative Program 8 Running the 24E Program 10 Running the RT Listeria P...

Page 4: ...e target temperature and instructs the module to heat or chill the plate as required The block automatically provides the sequential heating and cooling conditions for lysis protocols eliminating the...

Page 5: ...and proceed with appropriate caution Indicates the presence of a hot surface or other high temperature hazard and proceed with appropriate caution Indicates the presence of an electrical shock hazard...

Page 6: ...rmal block unit on a level dry laboratory workbench Allow at least 4 inches 10 cm on all sides of the unit for ventilation 2 Insert one end of the provided power cord into the power supply and plug th...

Page 7: ...display An audible beep sounds when the button is pressed 3 SELECT CONTINUE button sets the menu item in the LCD display as the selected program and is also used to confirm required actions An audible...

Page 8: ...mal block by toggling the on off switch on the back panel of the unit At start up the LCD automatically displays the following series of messages The first screen shows the version of software current...

Page 9: ...mples prompt appears place a rack of prepared samples into the metal tube holder Note See the BAX System User Guide for information on preparing samples Press the SELECT CONTINUE button 5 The LCD disp...

Page 10: ...e LCD display changes to Completed and the thermal block sounds an audible beep to signal that the program has been completed The timer continues to show any additional time that samples are held at 4...

Page 11: ...unit sounds two audible beeps and the blue ACTION NEEDED LED activates The LCD display changes to Load Samples 4 After the Load Samples prompt appears place a rack of prepared samples into the metal t...

Page 12: ...emoved Press the SELECT CONTINUE button Note A timer in the bottom right corner counts up from 0 00 to show any additional time that samples are held at 4 C 10 The LCD display changes to Completed and...

Page 13: ...ion Press the SELECT CONTINUE button 2 The unit beeps once and the LCD display changes as the thermal block begins equilibrating to 37 C Note Do not load samples into the thermal block until the load...

Page 14: ...ps once and the LCD display changes as the thermal block begins equilibrating to 55 C While the block is heating follow the instructions in the BAX System User Guide to add additional lysing reagents...

Page 15: ...he thermal block sounds two audible beeps to signal the change A timer in the bottom right corner counts up from 0 00 during the temperature change No user action is required 15 After samples have coo...

Page 16: ...dration 18 After the PCR tablets have been hydrated remove the rack of cluster tubes from the metal tube holder Press the SELECT CONTINUE button 19 The LCD display changes to 24E Finished and the ther...

Page 17: ...mples prompt appears place a rack of prepared samples into the metal tube holder Note See the BAX System User Guide for information on preparing samples Press the SELECT CONTINUE button 5 The LCD disp...

Page 18: ...0 The LCD display changes to Completed and the thermal block sounds an audible beep to signal that the program has been completed The timer continues to show any additional time that samples are held...

Page 19: ...rinse to prevent damage to the equipment Soak metal tube holders in 20 bleach solution for about 5 minutes then rinse thoroughly with water and allow to air dry An alternate cleaner designed to remove...

Page 20: ...value entered in the LCD display SELECT CONTINUE button decreases the value entered in the LCD display Pressing SCROLL and SELECT CONTINUE buttons at the same time sets the value in the LCD display a...

Page 21: ...g the on off switch on the back panel The LCD displays a message that Default Settings Are Now Loaded 6 Continue to hold the SELECT CONTINUE button until the LCD displays Set Point off about 15 second...

Page 22: ...ating process If needed use the SELECT CONTINUE button to decrease the value of the set point Note Be sure to wait for the temperature reading of the thermometer to equilibrate before proceeding 11 Pr...

Page 23: ...ured temperature If necessary the thermal block will automatically adjust the Plate Temperature to align with the Set Point 16 Press the SCROLL button to increase the value of the set point to 95 C Th...

Page 24: ...the SELECT CONTINUE button to decrease the value of the Measured temperature to the value displayed on the thermometer Note When entering the Measured temperature do not include decimal places 21 Pres...

Page 25: ...running press the SCROLL button to display the Abort menu 2 Press the SCROLL push button until the arrow is next to the Abort option then press the SELECT CONTINUE push button 3 The LCD display shows...

Page 26: ...ng any step in the program This fault indicator appears on the LCD display until the program completes or the program is aborted by the user Affected samples should not be used with the BAX System Rem...

Reviews: