20
Automated Thermal Block
13 Press the SELECT/CONTINUE push
button until the arrow
is next to the
Lo Temp Cal Pt
option.
Press the SCROLL and
SELECT/CONTINUE buttons at the
same time to select the
Lo Temp Cal
Pt
option.
14 The LCD display changes to a screen
that compares the Displayed and
Measured temperatures.
If needed, use the SCROLL button to
increase the value of the Measured
temperature or the
SELECT/CONTINUE button to
decrease the value of the Measured
temperature to the value displayed on
the thermometer.
Note: When entering the Measured
temperature, do not include decimal
places.
15 Press the SCROLL and
SELECT/CONTINUE buttons at the
same time to set the Measured
temperature.
If necessary, the thermal block will
automatically adjust the Plate
Temperature to align with the Set Point.
16 Press the SCROLL button to increase
the value of the set point to 95 °C. The
block automatically begins the heating
process.
If needed, use the
SELECT/CONTINUE button to
decrease the value of the set point.
Note: Be sure to wait for the temperature
reading of the thermometer to equilibrate
before proceeding
.
17 Press the SCROLL and
SELECT/CONTINUE buttons at the
same time. The calibration menu again
appears on the LCD display.
18 Press the SELECT/CONTINUE button
until the arrow
is next to the
Calibrate
option.
Press the SCROLL and
SELECT/CONTINUE buttons at the
same time to select the
Calibrate
option.
Lo Temp Cal Pt
Hi Temp Cal Pt
Displayed: 37C
Measured: 38C
Set Point: 37C
Plate Temp: 37C
Set Point: 95C
Plate Temp: 37C
Exit
Calibrate
Exit
Calibrate