User Guide
9
INS2051 Rev A
7
After the thermal block reaches 95 °C, the
LCD display changes to “Denaturing @ 95C”.
A timer in the bottom right corner counts down
from 10:00 and shows the number of minutes
remaining in this step. No user action is
required.
8
After denaturation at 95 °C is complete, the
LCD display changes to “Cooling to 4C” and
the thermal block sounds two audible beeps
to signal the change. A timer in the bottom
right corner counts up from 0:00 during the
temperature change. No user action is
required.
9
After samples have cooled for 5 minutes, the
LCD display changes to “Sample PCR
Ready”, the blue ACTION NEEDED LED
activates and the thermal block sounds four
audible beeps to signal that samples can be
removed.
Press the SELECT/CONTINUE button.
Note: A timer in the bottom right corner counts
up from 0:00 to show any additional time that
samples are held at 4 °C.
10 The LCD display changes to “Completed” and
the thermal block sounds an audible beep to
signal that the program has been completed.
The timer continues to show any additional
time that samples are held at 4 °C.
11 Follow the instructions in the BAX System
User Guide to hydrate the PCR tablets with
these lysates.
Note: Lysates must be maintained at 4 °C
during tablet hydration.
12 After the PCR tablets have been hydrated,
remove the rack of cluster tubes from the
metal tube holder.
Press the SELECT/CONTINUE button.
13 The LCD display changes to “Gram Negative
Finished” and the thermal block sounds an
audible beep to signal that the program has
finished.
Denaturing @95C
GmNeg 8:26
Cooling to 4C
GmNeg 2:15
Sample PCR Ready
GmNeg 1:01
Completed
GmNeg 4:54
Gram Negative
**Finished**
Summary of Contents for DUP-1000