User Guide
17
INS2051 Rev A
the lower resolution of the dry bath thermometer, calibration with an RTD thermometer is strongly
recommended.
The buttons on the front panel display will serve different purposes during calibration:
•
SCROLL
button – increases the value entered in the LCD display.
•
SELECT/CONTINUE
button – decreases the value entered in the LCD display.
•
Pressing
SCROLL
and
SELECT/CONTINUE
buttons at the same time - sets the value in
the LCD display as the selected value.
1
Arrange 48 cluster tubes in a rack and
fill each tube with 200 µL of water.
2
Poke a small hole in the cap of one
cluster tube near the center of the
cluster tube rack.
3
With the
power turned off
, place the
rack of cluster tubes in the metal
holder.
Summary of Contents for DUP-1000
Page 1: ...Model DUP 1000 User Guide...