18
Automated Thermal Block
4
Insert the microprobe of the RTD
thermometer through the hole in the
cap of the cluster tube.
Make sure the tip of the probe touches
the bottom of the tube to ensure that
the probe is completely immersed.
5
While pressing and holding the
SELECT/CONTINUE button, turn on
the power to the thermal block by
toggling the on/off switch on the back
panel. The LCD displays a message
that “Default Settings Are Now
Loaded”.
6
Continue to hold the
SELECT/CONTINUE button until the
LCD displays “Set Point: off” (about 15
seconds).
Note: Plate temp is not important at this
point.
7
Press the SCROLL and
SELECT/CONTINUE buttons at the
same time. The LCD displays the
Calibration menu.
8
Press the SELECT/CONTINUE button
until the arrow
is next to the
Clear
Cal Pts
option.
Default Settings
Are Now Loaded
Set Point: off
Plate Temp: 20C
Exit
Calibrate
Calibrate
Clear Cal Pts