User Guide
19
INS2051 Rev A
9
Press the SCROLL and
SELECT/CONTINUE push buttons at
the same time to select the
Clear Cal
Pts
option.
Calibration points previously saved in
the thermal block are cleared, and the
LCD display changes to read “Set
Point: off”.
10 Press the SCROLL button to increase
the value of the set point to 37 °C. The
block automatically begins the heating
process.
If needed, use the
SELECT/CONTINUE button to
decrease the value of the set point.
Note: Be sure to wait for the temperature
reading of the thermometer to equilibrate
before proceeding
.
11 Press the SCROLL and
SELECT/CONTINUE buttons at the
same time. The calibration menu again
appears on the LCD display.
12 Press the SELECT/CONTINUE button
until the arrow
is next to the
Calibrate
option.
Press the SCROLL and
SELECT/CONTINUE buttons at the
same time to select the
Calibrate
option.
Set Point: 20C
Plate Temp: 20C
Set Point: 37C
Plate Temp: 20C
Exit
Calibrate
Exit
Calibrate