8 After denaturation at 95 °C is complete, the LCD
display changes to “Cooling to 4C”, and the
thermal block sounds two audible beeps to signal
the change. A timer in the bottom right corner
counts up from 0:00 during the temperature
change. No user action is required.
9 After the samples have cooled for 5 minutes, the
LCD display changes to “Sample PCR Ready”, the
blue ACTION NEEDED LED activates and the
thermal block sounds four audible beeps to signal
that samples can be removed.
Press the SELECT/CONTINUE button.
Note: A timer in the bottom right corner
counts up from 0:00 to show any
additional time that samples are held at
4 °C.
10 The LCD display changes to “Completed” and the
thermal block sounds an audible beep to signal
that the program has been completed. The timer
continues to show any additional time that
samples are held at 4 °C.
11 Follow the instructions in the BAX System User
Guide to hydrate the PCR tablets with these
lysates.
Note: Lysates must be maintained at 4 °C
during tablet hydration.
12 After the PCR tablets have been hydrated, remove
the rack of cluster tubes from the metal tube
holder.
Press the SELECT/CONTINUE button.
13 The LCD display changes to “Gram Positive
Finished” and the thermal block sounds an audible
beep to signal that the program has finished.
14 Press the SELECT/CONTINUE button to exit from
the program. The LCD changes to display the
program menu.
Cooling to 4C
GmPos 1:45
Sample PCR Ready
GmPos 1:06
Completed
GmPos 6:25
Gram Positive
**Finished**