10
Automated Thermal Block
14 Press the SELECT/CONTINUE button to end
the program. The LCD changes to display the
program menu.
Running the 24E Program
Samples are heated at 37 °C for 30 minutes, then removed for the addition of lysis reagents.
Samples then are heated to 55 °C for 30 minutes, then 95 °C for 10 minutes, then held at 4 °C for
at least 5 minutes or until removed.
1
Turn on the thermal block, then press
the SCROLL button until the arrow
is
next to the
24E
option.
Press the SELECT/CONTINUE button.
2
The unit beeps once and the LCD
display changes as the thermal block
begins equilibrating to 37 °C.
Note: Do not load samples into the thermal
block until the load prompt appears on the
LCD display.
3
When the thermal block has reached
37 °C, the unit sounds two audible
beeps and the blue ACTION NEEDED
LED activates. The LCD display
changes to “Load Samples”.
4
After the “Load Samples” prompt
appears, place a rack of prepared
samples into the metal tube holder.
Note: See BAX System User Guide for more
information on prepared samples.
Press the SELECT/CONTINUE button.
Gram Negative
24E
24E
RT Listeria
Equilibr.@ 37C
24E
Load Samples
24E