Error messages / Troubleshooting
FluidAqua Mobil - FAM10/15-P
en(us)
Page 88 / 124
BEWA FAM10-15-P 4052078 en-us 2015-11-05.doc
2015-11-05
If several errors occur, then the most recently occurring message will be
displayed first. One can used the arrow keys
TAB
+
+/-
to scroll through the
individual fault messages.
All of the error messages shown in the display are equipped with a number. It
is very easy to find the error message again in this Chapter by using the
number as a reference.
To switch off the error messages, press the
button. If the fault indicator
lamp is flashing, error messages are present.
To show error messages again, switch to Menu 1 or Menu 2 and then return
to the display window.
1. CS+pump switched off. CS fault. Start 10 new attempts with Reset!
Cause:
The CS supplies 10 unusable measured values.
The CS and the charge pump are switched off for 15
seconds after each unsuccessful attempt, after which a
new measurement is carried out.
After an additional 10 unusable measured values, the
CS switches off and can be switched back on only by
actuating the Reset key.
The reasons for incorrect measurements could include:
no usable values or one of the errors 16, 17, 30, 31.
Remedy: Perform
the
following
steps:
Step
Description
1. Message:
1. CS-Zuförderp. Abgeschal. CS1xxxStörung. Start 10
new attempts with Reset!
2.
Actuate ESC key and reset failure with the F3 key.
3.
The measurement starts once again; automatic mode is not
interrupted.