Error messages / Troubleshooting
FluidAqua Mobil - FAM10/15-P
en(us)
Page 104 / 124
BEWA FAM10-15-P 4052078 en-us 2015-11-05.doc
2015-11-05
27. Fuse F2 has tripped
Cause:
The F2 fuse triggers. A message appears on the
display. The FAM switches off at once. The fault
indicator lamp is not flashing.
Remedy: Perform
the
following
steps:
Step
Description
1. Message:
27. Fuse F2 has tripped
2.
Switch the FAM off at the main switch and unplug the power
connector.
3.
Determine the cause and eliminate it.
4.
Switch on the unit at the main switch.
->a. The fault is repeated.
-> proceed to Step 2.
5.
->b. The fault is not repeated.
-> proceed to Step 6.
6.
The fault is repaired.