Error messages / Troubleshooting
FluidAqua Mobil - FAM10/15-P
en(us)
Page 100 / 124
BEWA FAM10-15-P 4052078 en-us 2015-11-05.doc
2015-11-05
24. Oil tank of rotary vane vacuum pump is empty -> refill oil
Cause:
The MIN switch point in the oil tank of the rotary vane
vacuum pump has been reached.
The message appears at once on the display.
The FAM goes into the after-running phase and
switches off.
Remedy: Perform
the
following
steps:
Step
Description
1. Message:
24. Oil tank of rotary vane vacuum pump is empty -> refill oil.
2.
Fill up the rotary vane vacuum pump with vacuum pump oil. For
details, see page 38.
3.
Actuate ESC key and reset failure with the F3 key.
4.
Start the unit by pressing the F1 key.
5.
The fault is repaired.
If the failure recurs, even though there is enough vacuum pump
oil between the MIN and MAX markings, first check the float
switch and afterwards the PLC.