AquaSensor - AS 8000 series
7.2.1 Parameter
Menu
Call up the parameter settings for the sensors in the display in figure 6 by pressing
<Strg><p>:
For Sensor 1:
Sensor Connected 0/1/2
0 when no sensor is connected.
1 when a sensor for the
range 5 to 80°C is connected.
2 when a sensor for the range
5 to 160°C is connected.
Alarm at % Comp-B
a value between 0 and 100 % is set, here 5%
20 mA at % Comp-B
the value for 4 mA is always 0% of component B [Comp B], or pure
component A.
The value for 20 mA is set here to 10%, so component A is
contaminated with 10% of component B.
Number of averaging (1-4)
The averaging of the measured values is done with
no averaging
/10-/30-/60-
measured values
For Sensor 2:
The following is displayed and set
Press <1>,
the values are written to the E-PROM of the Control Unit.
Press <2>,
all values can be changed Press <ESC>, return to main view 1. Figure 5
7.2.2 System constants menu
Call up the system constants menu from the Main Screen with the key combination <Strg>
<s>.
Fig. 7
To set the system constants on sensors 1 and 2, press the <1> or <2> keys for the
corresponding sensor. Press key <3> to call up the window for the basic settings of the 4-20
mA interface.
HYDAC Filtertechnik GmbH
en
Page 24
BeWa AS8000 3215331b en.doc
2007-07-12
Summary of Contents for AquaSensor AS 8000 Series