Commissioning
OPERATION MANUAL
Chapter 2
CC®-Immersion circulator
V2.4.0en/08.12.20//17.12
28
2.11.3
Converting the power supply connection
To restore the factory settings, the connection values must be
re-entered
.
The following temperature control units can be retrofitted:
CC-300BX, CC-315B(X)
▪
The temperature control units can be set to either 230 V/50 Hz or to 400 V/50 Hz.
▪
Install the temperature control unit at the location provided.
▪
The supply voltage must be selected during start-up for some temperature control units. This
selection can be changed later in the Pilot ONE menu ([Category menu] > [Installation] > [Power
supply]). The menu option “Voltage Supply” is visible only if the temperature control unit sup-
ports manually resetting the voltage supply. The supply voltage must be reselected if the temper-
ature control unit has been reset to the factory settings.