Setup mode
OPERATION MANUAL
Chapter 4
CC®-Immersion circulator
V2.4.0en/08.12.20//17.12
54
The correct selection will be displayed graphically and the “Max. Setpoint” will be changed
promptly. If tapping on “OK” is not correct, it will be displayed graphically for 2 seconds. The dis-
play will then return to the category “Setpoint Limits”. Try changing the “Max. Setpoint” again.
Check the set values for the minimum and maximum setpoint at any system change, especially
when changing the thermal fluid.
4.2.14
Setting the setpoint
PROCEDURE
Go to the “Home” screen.
Tap on the keypad symbol next to “T
Setpoint
”.
Enter a new setpoint using the number keypad that appears.
The following must apply:
[Minimum setpoint limit] ≤ [Setpoint] ≤ [Maximum setpoint limit].
If these conditions are not met, a message will appear on the
>Touchscreen<
[88] and the entry
will be ignored. In this case, delete the value just entered either with the “Arrow” key or with the
“clear” key. Enter the setpoint again.
Confirm your entry by tapping on “OK”.
In the display that follows, confirm your entry again by tapping on “OK”.
The correct selection will be displayed graphically and the setpoint will be changed immediately.
If tapping on “OK” is not correct, this is displayed graphically for 2 seconds. After this, the display
will return to the “Home” screen again. Try changing the setpoint again.
4.3
Filling, venting, degassing and draining
The illustration “connection diagram” can be found on page 73 in section
.
Extremely hot / cold surfaces, connections and thermal fluids
BURNS/FREEZING OF LIMBS
Surfaces, connections and tempered thermal fluids can be extremely hot or cold depending on
the operating mode.
Avoid direct contact with surfaces, connections and thermal fluids!
Wear your personnel protective equipment (e.g. temperature-resistant safety gloves, safety
goggles).
During an active circulation, the thermal fluid circuit is shut off by shut-off valves
MATERIAL DAMAGE TO THE CIRCULATING PUMPS INSTALLED IN THE TEMPERATURE CONTROL
UNIT
Do not close the thermal fluid circuit during an active circulation by means of shut-off valves.
Warm the thermal fluid to room temperature before stopping the circulation.
4.3.1
Filling, venting, degassing and draining the bath thermostat
Non-compliance with the safety data sheet for the thermal fluid to be used
INJURIES
Risk of injury to the eyes, skin, respiratory tract.
The safety data sheet for the thermal fluid to be used must be read prior to using it and its
content must be respected.
Observe the local regulations/work instructions.
Wear your personal protective equipment (e.g. temperature-resistant safety gloves, safety
goggles, safety footwear).
Danger of slipping because floor and work area are contaminated. Clean the work station and
follow the instructions for the disposal of thermal fluid and material on page 15 in section
»Proper disposal of resources and consumables«
.