Service/maintenance
Chapter 7
OPERATION MANUAL
V2.4.0en/08.12.20//17.12
CC®-Immersion circulator
65
7.4.2
Replacing temperature control or coolant hoses
Replace defective temperature control and/or coolant hoses
before
turning on the temperature
control unit.
7.4.2.1
Replacing temperature control hoses
PROCEDURE
Drain the temperature control unit as described on page 56 in Section
»Draining the bath thermostat«
.
Replace defective temperature control hoses. When disposing of them, observe on page 15
Section
»Proper disposal of resources and consumables«
.
Reconnect your external application as described on page 26 in Section
.
Fill the temperature control unit with thermal fluid as described on page 55 in Section
and venting the bath thermostat and the externally closed application«
.
Vent the temperature control unit as described on page 55 in Section
bath thermostat and the externally closed application«
.
Restart the temperature control unit in normal mode.
7.4.2.2
Replacing coolant hoses
PROCEDURE
Drain the cooling water as described on page 70 in section
.
Replace the defective coolant hoses. When disposing of them, observe on page 15 section
»Proper disposal of resources and consumables«
.
Reconnect the temperature control unit to the building’s cooling water supply. Proceed as de-
scribed on page 24 in section
»Temperature control units with cooler«
.
Restart the temperature control unit in normal mode.
7.5
Thermal fluid inspection, replacement and circuit cleaning
The illustration “connection diagram” can be found on page 73 in section
.
Extremely hot / cold surfaces, connections and thermal fluids
BURNS/FREEZING OF LIMBS
Surfaces, connections and tempered thermal fluids can be extremely hot or cold depending on
the operating mode.
Avoid direct contact with surfaces, connections and thermal fluids!
Wear your personnel protective equipment (e.g. temperature-resistant safety gloves, safety
goggles).
7.5.1
Thermal fluid inspection
Thermal fluid is not inspected on a regular basis
BURNS DUE TO REDUCED BOILING POINT
Regularly check your thermal fluid whether it meets the specifications in the safety data sheet.
Thermal fluid is not inspected on a regular basis
DAMAGE TO THE HEAT EXCHANGER AND/OR ELECTROMECHANICAL PARTS.
Regularly check your thermal fluid whether it meets the specifications in the safety data sheet.