Setup mode
Chapter 4
OPERATION MANUAL
V2.4.0en/08.12.20//17.12
CC®-Immersion circulator
57
PROCEDURE
Baths with >Drain valve< [4]
Remove the knurled screw at the
>Drain<
[8].
Connect a suitable drain hose to the
>Drain<
[8].
Place the other end of the hose in a suitable container.
Open the
>Drain valve<
[4]
by turning it counterclockwise (turn 90° left as far as it will go). The
thermofluid will flow from the external application via the bath vessel and the draining hose into
the container. Check if the thermofluid can be reused. On page 15 observe section
posal of resources and consumables«
.
Wait until the external application and the bath are empty.
Open the connection
>Circulation flow<
[1].
Open the connection
>Circulation return<
[2].
Leave the temperature control unit open for a while to allow it to dry out and the residue to
drain. Without screw caps and with the
>Drain valve<
[4] open.
Close the
>Drain valve<
[4]
by turning it clockwise (turn 90° right as far as it will g
o).
Close the connection
>Circulation flow<
[1].
Close the connection
>Circulation return<
[2].
After drying out, remove the drain hose and re-fit the knurled screw to the
>Drain<
[8].
The bath is now drained.
Baths without >Drain valve< [4]
Have a suitable container ready to catch the thermal fluid.
Open the knurled screw at the
>Drain<
[8]. As soon as you have opened the knurled screw, the
thermal fluid will flow from the external application over the bath and into the container. Wait
until the external application and the bath are empty. Check if the thermal fluid can be reused.
On page 15 observe Section
»Proper disposal of resources and consumables«
.
Open the connection
>Circulation flow<
[1].
Open the connection
>Circulation return<
[2].
Leave the temperature control unit open for a while for the residue to fully drain and to allow it
to dry out (without screw caps).
Close the connection
>Circulation flow<
[1].
Close the connection
>Circulation return<
[2].
Re-fit the knurled screw to the
>Drain<
[8].
The bath is now drained.