Commissioning
Chapter 2
OPERATION MANUAL
V2.4.0en/08.12.20//17.12
CC®-Immersion circulator
21
Use of the temperature control unit is permitted only under normal ambient conditions in accord-
ance with DIN EN 61010-1:2011:
▪
Use only indoors. The illuminance must be at least 300 lx.
▪
Installation altitude up to 2000 meters above sea level .
▪
Maintain wall and ceiling clearance for adequate air exchange (dissipation of waste heat, supply
of fresh air for the temperature control unit and work area). Ensure adequate floor clearance for
air-cooled temperature control units. Do not operate this temperature control unit from within
the box or with an inadequately dimensioned bath. This inhibits the exchange of air.
▪
Ambient temperature values are provided on the technical data sheet; compliance with the ambi-
ent conditions is mandatory, to ensure trouble-free operation.
▪
Relative humidity up to 32 °C max. 80% and decreasing linearly to 50% up to 40 °C.
▪
Short distance to supply connections.
▪
The temperature control unit must not be installed so as to hinder or prevent access to the isola-
tor (to the power supply).
▪
Magnitude of the power supply fluctuations: see data sheet from page 73 in section
▪
Transient surges, as would normally occur in the power supply system
▪
Installation Class 3
▪
Applicable pollution degree: 2.
▪
Surge category II.
Side
Distance in cm
[A2]
Top
can be located under a bench
[B]
Left
min. 10
[C]
Right
min. 10
[D]
Front
min. 10
[E]
Rear
min. 10
Side
Distance in cm (for operation in a tub)
[A2]
Top
can be located under a bench
[B]
Left
min. 20
[C]
Right
min. 20
[D]
Front
min. 20
[E]
Rear
min. 20
Wall clearances