Chapter 9
249
Performance Verification Tests
13. Frequency Response
21.Adjust the frequency synthesizer amplitude until the spectrum
analyzer marker reads
−
14 dBm. This corresponds to the amplitude
at 100 kHz recorded in
step 10
. Record the DVM amplitude in
column 2 of
Table 9-16
for DVM Amplitude at 100 kHz.
22.On the spectrum analyzer, press
Peak Search
,
Marker
,
Marker
∆
.
23.Set the spectrum analyzer and the frequency synthesizer to the next
frequency settings listed in
Table 9-16
.
24.At each frequency, adjust the frequency synthesizer amplitude for a
Sig-
∆
-Trk amplitude reading of 0.00
±
0.05 dB.
25.Record the DVM amplitude setting in column 2 of
Table 9-16
as the
DVM amplitude.
26.Repeat
step 23
through
step 25
for each frequency setting listed in
Table 9-16
.
27.For each of the frequencies in
Table 9-16
, subtract the DVM
Amplitude (column 2) from the DVM Amplitude at 100 kHz recorded
in
step 21
. Record the result as the Response Relative to 100 kHz
(column 3) of
Table 9-16
.
28.Add the entry from
Table 9-15
column 2 (100 kHz Error Relative to
50 MHz) that was recorded in
step 10
, to each of the entries in
Table
9-16
column 3 (Response Relative to 100 kHz). Record the results as
the Response Relative to 50 MHz (column 4) in
Table 9-16
.
Test Results
Perform the following steps to verify the frequency response of the
spectrum analyzer.
1. Enter the most positive number from
Table 9-16
, column 4:
____________ dB
If your analyzer is equipped with 75
Ω
Input, no entry is necessary.
2. Enter the most positive number from
Table 9-15
, column 2:
____________ dB For 75
Ω
Input use column 5.
3. Record the more positive of numbers from
step 1
and
step 2
in TR
Entry 1 of the performance verification test record.
4. Enter the most negative number from
Table 9-16
, column 4:
____________ dB
If your analyzer is equipped with 75
Ω
Input, no entry is necessary.
5. Enter the most negative number from
Table 9-15
, column 2:
____________ dB For 75
Ω
Input use column 5.
Summary of Contents for ESA-L1500A
Page 12: ...11 1 Preparing for Use and Printing ...
Page 28: ...27 2 Getting Started ...
Page 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Page 48: ...47 3 Troubleshooting ...
Page 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Page 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Page 118: ...117 5 Menu Maps ...
Page 119: ...118 Chapter5 Menu Maps ...
Page 120: ...Chapter 5 119 Menu Maps ...
Page 121: ...120 Chapter5 Menu Maps ...
Page 122: ...Chapter 5 121 Menu Maps ...
Page 124: ...Chapter 5 123 Menu Maps ...
Page 125: ...124 Chapter5 Menu Maps ...
Page 128: ...Chapter 5 127 Menu Maps ...
Page 129: ...128 Chapter5 Menu Maps ...
Page 130: ...Chapter 5 129 Menu Maps ...
Page 131: ...130 Chapter5 Menu Maps ...
Page 135: ...134 Chapter5 Menu Maps ...
Page 136: ...135 6 Options and Accessories ...
Page 145: ...144 Chapter6 Options and Accessories Accessories ...
Page 146: ...145 7 Specifications ...
Page 179: ...178 Chapter7 Specifications Regulatory Information ...
Page 180: ...179 8 Required Equipment ...
Page 187: ...186 Chapter8 Required Equipment Test Equipment ...
Page 308: ...307 10 Performance Verification Test Record ...