Chapter 4
91
Front-Panel Key Reference
Measure
N dB Points On Off
automatically places two markers at points N dB from the highest point
on the highest displayed signal, and determines the frequency
difference between the two markers. N dB is the active function and the
value of N is set by the user. The measurement defaults to -3 dB when it
is first turned on. The measurement runs continuously, re-executing at
the end of each sweep.
No other signal can appear on the display within N dB of the highest
signal. The measured signal cannot have more than one peak that is
greater than or equal to N dB. A signal must be greater than the peak
excursion above the threshold to be identified. The setting for peak
excursion may be increased from the 6 dB default value so that noise
will not be identified as signals. Increasing the value too much may
cause a smaller signal to be missed or misinterpreted as part of a larger
signal. The amplitude scale may be either linear or logarithmic.
Key Access:
Measure
% AM On Off
determines the percent of amplitude modulation of the largest
displayed signal and its two sidebands. The sidebands are assumed to
be entirely from amplitude modulation. If there are differences in the
sideband amplitude, the larger value is used. The measurement runs
continuously, re-executing at the end of each sweep.
The three signals that make up the AM signal, must all be displayed.
The frequency spacing of the sideband signals must be the same within
the span accuracy of the measurement. All of the signals must be
greater than the peak excursion above the threshold. The amplitude
scale may be either linear or logarithmic. Key Access:
Measure
TOI On Off
finds the third-order intercept of the two highest amplitude signals and
the two associated distortion products. The measurement compensates
for unequal test signal amplitude. The measurement runs continuously,
re-executing at the end of each sweep. The units for the displayed value
can be selected by pressing the
Amptd Units
softkey. The two test signals
and the two associated distortion products must all be displayed for the
measurement to function. The relative amplitudes and frequencies of
the displayed signals must fit the TOI pattern. All of the signals must
be greater than the peak excursion above the threshold.
Key Access:
Measure
Summary of Contents for ESA-L1500A
Page 12: ...11 1 Preparing for Use and Printing ...
Page 28: ...27 2 Getting Started ...
Page 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Page 48: ...47 3 Troubleshooting ...
Page 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Page 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Page 118: ...117 5 Menu Maps ...
Page 119: ...118 Chapter5 Menu Maps ...
Page 120: ...Chapter 5 119 Menu Maps ...
Page 121: ...120 Chapter5 Menu Maps ...
Page 122: ...Chapter 5 121 Menu Maps ...
Page 124: ...Chapter 5 123 Menu Maps ...
Page 125: ...124 Chapter5 Menu Maps ...
Page 128: ...Chapter 5 127 Menu Maps ...
Page 129: ...128 Chapter5 Menu Maps ...
Page 130: ...Chapter 5 129 Menu Maps ...
Page 131: ...130 Chapter5 Menu Maps ...
Page 135: ...134 Chapter5 Menu Maps ...
Page 136: ...135 6 Options and Accessories ...
Page 145: ...144 Chapter6 Options and Accessories Accessories ...
Page 146: ...145 7 Specifications ...
Page 179: ...178 Chapter7 Specifications Regulatory Information ...
Page 180: ...179 8 Required Equipment ...
Page 187: ...186 Chapter8 Required Equipment Test Equipment ...
Page 308: ...307 10 Performance Verification Test Record ...