Chapter 9
283
Performance Verification Tests
20. Tracking Generator Feedthrough Options IDN and IDQ
20. Tracking Generator Feedthrough
Options IDN and IDQ
The tracking generator output is terminated and set for 0 dBm output
power (maximum output power). The analyzer input is also terminated.
The noise level of the analyzer is then measured at several frequencies.
There are no related adjustments for this performance test.
Equipment Required
Termination, 50
Ω
, Type-N (m) (2 required)
Additional Equipment Required For 75
Ω
RF Out
Termination, 75
Ω
,Type-N (m), (2 required)
Adapter, Type-N (m) to BNC (m), (2 required)
CAUTION
Use only 75
Ω
cables, connectors, or adapters on instruments with 75
Ω
connectors or damage to the connectors will occur.
Figure 9-23
Tracking Generator Feedthrough Test Setup
Procedure
1. Press
Preset
on the analyzer, then wait for the preset routine to
finish. Set the analyzer by pressing the following keys:
System
,
50 MHz osc On Off
(On)
Frequency
,
50 MHz
Span
,
2 kHz
Summary of Contents for ESA-L1500A
Page 12: ...11 1 Preparing for Use and Printing ...
Page 28: ...27 2 Getting Started ...
Page 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Page 48: ...47 3 Troubleshooting ...
Page 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Page 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Page 118: ...117 5 Menu Maps ...
Page 119: ...118 Chapter5 Menu Maps ...
Page 120: ...Chapter 5 119 Menu Maps ...
Page 121: ...120 Chapter5 Menu Maps ...
Page 122: ...Chapter 5 121 Menu Maps ...
Page 124: ...Chapter 5 123 Menu Maps ...
Page 125: ...124 Chapter5 Menu Maps ...
Page 128: ...Chapter 5 127 Menu Maps ...
Page 129: ...128 Chapter5 Menu Maps ...
Page 130: ...Chapter 5 129 Menu Maps ...
Page 131: ...130 Chapter5 Menu Maps ...
Page 135: ...134 Chapter5 Menu Maps ...
Page 136: ...135 6 Options and Accessories ...
Page 145: ...144 Chapter6 Options and Accessories Accessories ...
Page 146: ...145 7 Specifications ...
Page 179: ...178 Chapter7 Specifications Regulatory Information ...
Page 180: ...179 8 Required Equipment ...
Page 187: ...186 Chapter8 Required Equipment Test Equipment ...
Page 308: ...307 10 Performance Verification Test Record ...