248
Chapter 9
Performance Verification Tests
13. Frequency Response
Figure 9-14
Frequency Response Test Setup, <100 kHz
Frequency Response, < 100 kHz
NOTE
If your analyzer is equipped with 75
Ω
Input, skip this section and go to
Test Results
.
18.Connect the equipment as shown in
Figure 9-14
.
Set the frequency synthesizer controls as follows:
FREQUENCY
,
100 kHz
AMPLITUDE
,
−
8 dBm
AMPTD INCR
,
0.05 dB
19.Set the DVM as follows:
Function
Sync AC Volts
Math
dBm
RES Register
50
Ω
Front/Rear Terminals
Front
Resolution
7.5 digits
20.On the analyzer, press the following keys:
Frequency
,
100 kHz
Summary of Contents for ESA-L1500A
Page 12: ...11 1 Preparing for Use and Printing ...
Page 28: ...27 2 Getting Started ...
Page 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Page 48: ...47 3 Troubleshooting ...
Page 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Page 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Page 118: ...117 5 Menu Maps ...
Page 119: ...118 Chapter5 Menu Maps ...
Page 120: ...Chapter 5 119 Menu Maps ...
Page 121: ...120 Chapter5 Menu Maps ...
Page 122: ...Chapter 5 121 Menu Maps ...
Page 124: ...Chapter 5 123 Menu Maps ...
Page 125: ...124 Chapter5 Menu Maps ...
Page 128: ...Chapter 5 127 Menu Maps ...
Page 129: ...128 Chapter5 Menu Maps ...
Page 130: ...Chapter 5 129 Menu Maps ...
Page 131: ...130 Chapter5 Menu Maps ...
Page 135: ...134 Chapter5 Menu Maps ...
Page 136: ...135 6 Options and Accessories ...
Page 145: ...144 Chapter6 Options and Accessories Accessories ...
Page 146: ...145 7 Specifications ...
Page 179: ...178 Chapter7 Specifications Regulatory Information ...
Page 180: ...179 8 Required Equipment ...
Page 187: ...186 Chapter8 Required Equipment Test Equipment ...
Page 308: ...307 10 Performance Verification Test Record ...