Performance Verification Tests
10. Resolution Bandwidth Accuracy
Figure 9-9
Resolution Bandwidth Accuracy Test Setup
CAUTION
Use only 75
Ω
cables, connectors, or adapters on instruments with 75
Ω
connectors, or damage to the connectors will occur.
Procedure
1. Connect the equipment as shown in
Figure 9-9
.
3 dB Bandwidths
2. Set the synthesized signal generator controls as follows:
AMPLITUDE
,
0 dBm
(50
Ω
Input only)
AMPLITUDE
,
6 dBm (75
Ω
Input only)
FREQUENCY
,
50 MHz
3. Press
Preset
on the analyzer, then wait for the preset routine to
finish. Press
System, Alignments, Auto Align, Off
. Set the analyzer by
pressing the following keys:
Frequency
,
50 MHz
Span
,
4.5 MHz
BW/Avg
,
3 MHz
Video BW Auto Man
,
30 Hz
Amplitude
,
Scale/Div 1 dB
Summary of Contents for ESA-L1500A
Page 12: ...11 1 Preparing for Use and Printing ...
Page 28: ...27 2 Getting Started ...
Page 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Page 48: ...47 3 Troubleshooting ...
Page 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Page 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Page 118: ...117 5 Menu Maps ...
Page 119: ...118 Chapter5 Menu Maps ...
Page 120: ...Chapter 5 119 Menu Maps ...
Page 121: ...120 Chapter5 Menu Maps ...
Page 122: ...Chapter 5 121 Menu Maps ...
Page 124: ...Chapter 5 123 Menu Maps ...
Page 125: ...124 Chapter5 Menu Maps ...
Page 128: ...Chapter 5 127 Menu Maps ...
Page 129: ...128 Chapter5 Menu Maps ...
Page 130: ...Chapter 5 129 Menu Maps ...
Page 131: ...130 Chapter5 Menu Maps ...
Page 135: ...134 Chapter5 Menu Maps ...
Page 136: ...135 6 Options and Accessories ...
Page 145: ...144 Chapter6 Options and Accessories Accessories ...
Page 146: ...145 7 Specifications ...
Page 179: ...178 Chapter7 Specifications Regulatory Information ...
Page 180: ...179 8 Required Equipment ...
Page 187: ...186 Chapter8 Required Equipment Test Equipment ...
Page 308: ...307 10 Performance Verification Test Record ...