Chapter 9
267
Performance Verification Tests
16. Other Input Related Spurious Responses
Peak Search
Marker, Marker Normal
Search, Mkr
→
Ref Lvl
Peak Search
Marker, Marker
∆
Single Sweep
24.On the synthesized sweeper, press
RECALL 3
for a CW frequency of
1310.7 MHz
25.Press
Single Sweep
on the analyzer and wait for a completion of a
new sweep.
26.On the analyzer, press
Peak Search
and record the marker delta
amplitude in the performance test record as TR Entry 3.
Summary of Contents for ESA-L1500A
Page 12: ...11 1 Preparing for Use and Printing ...
Page 28: ...27 2 Getting Started ...
Page 47: ...46 Chapter2 Getting Started Analyzer Battery Information ...
Page 48: ...47 3 Troubleshooting ...
Page 54: ...Chapter 3 53 Troubleshooting How to Return Your Analyzer for Service Figure 3 1 ...
Page 117: ...116 Chapter4 Front Panel Key Reference Trig ...
Page 118: ...117 5 Menu Maps ...
Page 119: ...118 Chapter5 Menu Maps ...
Page 120: ...Chapter 5 119 Menu Maps ...
Page 121: ...120 Chapter5 Menu Maps ...
Page 122: ...Chapter 5 121 Menu Maps ...
Page 124: ...Chapter 5 123 Menu Maps ...
Page 125: ...124 Chapter5 Menu Maps ...
Page 128: ...Chapter 5 127 Menu Maps ...
Page 129: ...128 Chapter5 Menu Maps ...
Page 130: ...Chapter 5 129 Menu Maps ...
Page 131: ...130 Chapter5 Menu Maps ...
Page 135: ...134 Chapter5 Menu Maps ...
Page 136: ...135 6 Options and Accessories ...
Page 145: ...144 Chapter6 Options and Accessories Accessories ...
Page 146: ...145 7 Specifications ...
Page 179: ...178 Chapter7 Specifications Regulatory Information ...
Page 180: ...179 8 Required Equipment ...
Page 187: ...186 Chapter8 Required Equipment Test Equipment ...
Page 308: ...307 10 Performance Verification Test Record ...