Sweeping
The AFG offers linear frequency sweeping of standard waveforms
(i.e. sine, square, triangle, ramp) and arbitrary waveforms from 0.0 Hz to
15 MHz for sine waves, and 0.0 Hz to 1 MHz for square, triangle, and ramp
waves.
The command subsystems covered in this section include:
•
[SOURce:]ROSCillator
•
[SOURce:]FREQuency
•
[SOURce:]SWEep
Sweeping
Considerations
The AFG only allows for continuous repetitive sweeping. The
INITiate
[:IMMediate]
command starts the sweeping at a fixed rate determined by the
number of points in the sweep and the sweep time. For more information on
sweeping, see Chapter 7 under the
[SOURce:]SWEep
subsystem.
Sweeping Using
Start and Stop
Frequencies
The following program specifies a start frequency and a stop frequency and
continuously sweeps between 0 and 1 MHz. The program also queries the
start frequency, stop frequency, center frequency, and frequency span to
show the relationship between them.
Using the flowchart as a guide, the steps of this program are:
1. Select the Reference Oscillator -
[SOURce:]ROSCillator:SOURce <
source
>
2. Select the Frequency Sweep Mode -
[SOURce:]FREQuency:MODE SWEep
3. Set the Start Frequency -
[SOURce:]FREQuency:STARt <
start_freq
>
4. Set the Stop Frequency -
[SOURce:]FREQuency:STOP <
stop_freq
>
5. Select the Output Function -
[SOURce:]FUNCtion[:SHAPe] <
shape
>
6. Set the Signal Amplitude -
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <
amplitude
>
7. Place the AFG in the Wait-for-arm State -
INITiate[:IMMediate]
85 HP E1340A Sweeping and Frequency-Shift Keying
Chapter 4
Summary of Contents for E1340A
Page 12: ...Notes 12 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 14: ...14 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 42: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 42 ...
Page 54: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 54 ...
Page 58: ...Chapter 3 Generating Arbitrary Waveforms with the HP E1340A 58 ...
Page 84: ...Chapter 4 HP E1340A Sweeping and Frequency Shift Keying 84 ...
Page 130: ...Chapter 6 HP E1340A High Speed Operation 130 ...
Page 202: ...202 HP E1340A SCPI Conformance Information Chapter 7 ...
Page 218: ...218 HP E1340A Specifications Appendix A ...
Page 284: ...284 HP E1340A Register Based Programming Appendix C ...
Page 295: ...Index HP E1340A Arbitrary Function Generator User s Manual 295 ...