[SOURce:]RAMP
The
[SOURce:]RAMP
subsystem selects the polarity of ramp waveforms.
Subsystem Syntax
[SOURce:]
RAMP
:POLarity <
polarity
>
:POLarity
[SOURce:]RAMP:POLarity <
polarity
>
selects the polarity of the ramp
waveform.
NORMal
generates a positive-going ramp;
INVerted
generates a
negative-going ramp.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
polarity
discrete
INVerted | NORMal
none
Comments
•
Executable when initiated: Query form only
•
Coupled command: Yes
•
Related Commands:
SOURce:FUNCtion:SHAPe
•
*RST Condition:
SOURce:RAMP:POLarity NORMal
Example
Selecting ramp polarity
FUNC:SHAP RAMP
Select ramp output
RAMP:POL INV
Select negative-going ramp
[SOURce:]RAMP:POLarity
Chapter 7
HP E1340A Command Reference 169
Summary of Contents for E1340A
Page 12: ...Notes 12 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 14: ...14 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 42: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 42 ...
Page 54: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 54 ...
Page 58: ...Chapter 3 Generating Arbitrary Waveforms with the HP E1340A 58 ...
Page 84: ...Chapter 4 HP E1340A Sweeping and Frequency Shift Keying 84 ...
Page 130: ...Chapter 6 HP E1340A High Speed Operation 130 ...
Page 202: ...202 HP E1340A SCPI Conformance Information Chapter 7 ...
Page 218: ...218 HP E1340A Specifications Appendix A ...
Page 284: ...284 HP E1340A Register Based Programming Appendix C ...
Page 295: ...Index HP E1340A Arbitrary Function Generator User s Manual 295 ...