:LOAD
OUTPut:LOAD <
load
>
indicates whether the actual load applied to the HP
E1340A’s "Aux Out" BNC is either matched to the output impedance
specified by
OUTPut:IMPedance
or is an open circuit. The output voltage
into an open circuit is twice that into a matched load. Setting
OUTPut:LOAD INFinity
compensates for this effect so that the
SOURce:LIST:SEGMent:VOLTage
and
SOURce:VOLTage:LEVel:IMMediate:AMPLitude
and
OFFSet
commands
will output the specified voltages into an open circuit.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
load
numeric
50 |
9.9E+37 | INFinity |
MINimum | MAXimum
Ohms
MINimum and MAXimum both select 50
Ω
load impedance.
Use 9.9E+37 or INFinity to indicate an open circuit output.
Comments
•
Executable when initiated: Query form only
•
Coupled command: Yes
•
Related commands:
OUTPut:IMPedance
,
SOURce:LIST
subsystem,
SOURce:VOLTage
subsystem
•
*RST Condition:
OUTPut:LOAD 50
Example
Indicating open circuit output load
OUTP:LOAD INF
Indicate open circuit
OUTPut:LOAD
148 HP E1340A Command Reference
Chapter 7
Summary of Contents for E1340A
Page 12: ...Notes 12 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 14: ...14 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 42: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 42 ...
Page 54: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 54 ...
Page 58: ...Chapter 3 Generating Arbitrary Waveforms with the HP E1340A 58 ...
Page 84: ...Chapter 4 HP E1340A Sweeping and Frequency Shift Keying 84 ...
Page 130: ...Chapter 6 HP E1340A High Speed Operation 130 ...
Page 202: ...202 HP E1340A SCPI Conformance Information Chapter 7 ...
Page 218: ...218 HP E1340A Specifications Appendix A ...
Page 284: ...284 HP E1340A Register Based Programming Appendix C ...
Page 295: ...Index HP E1340A Arbitrary Function Generator User s Manual 295 ...