:POINts
[SOURce:]SWEep:POINts <
number
>
selects the number of points in a
frequency sweep.
The frequencies generated by the sweep are evenly spaced linearly between
the frequencies specified by
SOURce:FREQuency:STARt
and
STOP
, or
CENTer
and
SPAN
, inclusive.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
number
numeric
see below |
MINimum | MAXimum
none
The legal range for
number
, as well as the MINimum and MAXimum values, is
context-dependent. See the section “Sweep Considerations” at the beginning
of the SOURce:SWEep subsystem for a description of the dependencies.
Comments
•
Executable when initiated: Query form only
•
Coupled command: Yes
•
Related Commands:
SOURce:FREQuency:CENTer
,
MODE
,
SPAN
,
START
, and
STOP
•
*RST Condition:
SOURce:SWEep:POINts 1001
Example
Setting the number of points in the sweep
SWE:POIN 100
Set 100 points in sweep
:TIME
[SOURce:]SWEep:TIME <
time
>
selects the duration of the sweep. The
duration is the time from the start of the sweep until when the last frequency
begins to be output.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
time
numeric
see below |
MINimum | MAXimum
seconds
With CALibration:STATe:AC OFF set: MINimum selects 0.187 mS *
(
points
- 1); MAXimum selects 25.687 mS * (
points
- 1).
With CALibration:STATe:AC ON set: MINimum selects 0.250 mS *
(
points
- 1); MAXimum selects 25.75 mS * (
points
- 1).
points
is specified by the SOURce:SWEep:POINts command.
The above values bound the valid range for
time
.
[SOURce:]SWEep:POINts
Chapter 7
HP E1340A Command Reference 175
Summary of Contents for E1340A
Page 12: ...Notes 12 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 14: ...14 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 42: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 42 ...
Page 54: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 54 ...
Page 58: ...Chapter 3 Generating Arbitrary Waveforms with the HP E1340A 58 ...
Page 84: ...Chapter 4 HP E1340A Sweeping and Frequency Shift Keying 84 ...
Page 130: ...Chapter 6 HP E1340A High Speed Operation 130 ...
Page 202: ...202 HP E1340A SCPI Conformance Information Chapter 7 ...
Page 218: ...218 HP E1340A Specifications Appendix A ...
Page 284: ...284 HP E1340A Register Based Programming Appendix C ...
Page 295: ...Index HP E1340A Arbitrary Function Generator User s Manual 295 ...