OUTPut
The
OUTPut
subsystem controls the characteristics of the output waveform.
The subsystem sets the output source impedance, and enables or disables
the output.
Subsystem Syntax
OUTPut
:IMPedance <
impedance
>
:LOAD <
load
>
:IMPedance
OUTPut:IMPedance <
impedance
>
sets the HP E1340A’s output
impedance.
Since the HP E1340A only provides 50
Ω
output impedance, there is no
need to send this command. It is provided for SCPI compatibility only.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
impedance
numeric
50
MINimum | MAXimum
Ohms
MINimum and MAXimum select 50
Ω
output impedance.
Comments
•
Impedance values between 25 and 62.5 will be accepted and rounded
to 50
Ω
. Values outside that range will cause error -222,"Argument
out of range" to occur.
•
Executable when initiated: Yes
•
Coupled command: Yes
•
Related commands:
OUTPut:LOAD
•
*RST Condition:
OUTPut:IMPedance 50
Example
Querying output impedance
OUTP:IMP?
50 will be returned with this query
OUTPut:IMPedance
Chapter 7
HP E1340A Command Reference 147
Summary of Contents for E1340A
Page 12: ...Notes 12 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 14: ...14 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 42: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 42 ...
Page 54: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 54 ...
Page 58: ...Chapter 3 Generating Arbitrary Waveforms with the HP E1340A 58 ...
Page 84: ...Chapter 4 HP E1340A Sweeping and Frequency Shift Keying 84 ...
Page 130: ...Chapter 6 HP E1340A High Speed Operation 130 ...
Page 202: ...202 HP E1340A SCPI Conformance Information Chapter 7 ...
Page 218: ...218 HP E1340A Specifications Appendix A ...
Page 284: ...284 HP E1340A Register Based Programming Appendix C ...
Page 295: ...Index HP E1340A Arbitrary Function Generator User s Manual 295 ...