6. Select the Waveform Segment -
[SOURce:]LIST[:SEGMent]:SELect <
name
>
Use either A, B, C, or D in
<
name
>
for the waveform segment.
7. Store the Waveform Segment as Voltages -
[SOURce:]LIST[:SEGMent]:VOLTage <
voltage_list
>
This command stores the points of the waveform segment into the
AFG’s segment memory. These points are sent to the AFG as volts
which are the output voltage points that constitutes the waveform
segment.
8. Select the User Name -
[SOURce:]FUNCtion:USER <
name
>
This command selects the waveform segment to be output. Make the
name in this command the same name as the waveform segment to be
output.
9. Initiate the Waveform -
INITiate[:IMMediate]
This command generates an immediate output with the arm source
set to
IMMediate
. Refer to Chapter 5 for triggering information.
10. Query the Waveform Segment (Optional) -
[SOURce:]LIST[:SEGment]:SELect?
This command returns the currently selected waveform segment.
Chapter 3
Generating Arbitrary Waveforms with the HP E1340A 60
Summary of Contents for E1340A
Page 12: ...Notes 12 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 14: ...14 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 42: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 42 ...
Page 54: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 54 ...
Page 58: ...Chapter 3 Generating Arbitrary Waveforms with the HP E1340A 58 ...
Page 84: ...Chapter 4 HP E1340A Sweeping and Frequency Shift Keying 84 ...
Page 130: ...Chapter 6 HP E1340A High Speed Operation 130 ...
Page 202: ...202 HP E1340A SCPI Conformance Information Chapter 7 ...
Page 218: ...218 HP E1340A Specifications Appendix A ...
Page 284: ...284 HP E1340A Register Based Programming Appendix C ...
Page 295: ...Index HP E1340A Arbitrary Function Generator User s Manual 295 ...