Program Example
This program shows how to store a waveform segment (i.e., points of an
arbitrary waveform) into the AFG’s segment memory. The waveform
segment are sent as DAC codes using the Definite Length Block Data
transfer method. The example generates a 4096 point 0 to +5 V positive
going ramp.
1. Reset the AFG -
*RST
2. Setup the AFG for Output -
[SOURce:]FREQuency[:CW|FIXed] <
frequency
>
[SOURce:]FUNCtion[:SHAPe] USER
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <
amplitude
>
3. Select the DAC Data Source -
[SOURce:]ARBitrary:DAC:SOURce INTernal
This command selects the source that transfers data to the DAC (see
“ Program Comments” ). Use
INTernal
to transfer the data using the
[SOURce:]LIST[1]
subsystem.
4. Select the Waveform Segment -
[SOURce:]LIST[:SEGMent]:SELect <
name
>
Use either
A
,
B
,
C
, or
D
in
<
name
>
for the waveform segment.
5. Store the Waveform Segment as DAC Data -
[SOURce:]LIST[:SEGMent]:VOLTage:DAC <
voltage_list
>
This command stores the waveform segment into segment memory
using DAC codes sent as Definite Length Block Data.
6. Generate the Output -
[SOURce:]FUNCtion:USER <
name
>
INITiate[:IMMediate]
Chapter 6
HP E1340A High Speed Operation 114
Summary of Contents for E1340A
Page 12: ...Notes 12 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 14: ...14 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 42: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 42 ...
Page 54: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 54 ...
Page 58: ...Chapter 3 Generating Arbitrary Waveforms with the HP E1340A 58 ...
Page 84: ...Chapter 4 HP E1340A Sweeping and Frequency Shift Keying 84 ...
Page 130: ...Chapter 6 HP E1340A High Speed Operation 130 ...
Page 202: ...202 HP E1340A SCPI Conformance Information Chapter 7 ...
Page 218: ...218 HP E1340A Specifications Appendix A ...
Page 284: ...284 HP E1340A Register Based Programming Appendix C ...
Page 295: ...Index HP E1340A Arbitrary Function Generator User s Manual 295 ...