Setting the function and loading RAM
Command Opcode = 5
Bit
7
6
5
4
3
2
1
0
Bit
Weight
128
64
32
16
8
4
2
1
Purpose
DAC Source
Download Mode
Waveform selection (waveform EEPROM)
RAM
Setting
0 - Input Data
Register
1 - waveform
EEPROM
0 - waveform
appears at output
as it is loaded into
RAM from the Input
Data Register.
1 - output remains
at 0V as waveform
is loaded into RAM
from the Input Data
Register.
0 0 0 0 - Sine wave
0 0 0 1 - Triangle wave
0 0 1 0 - Sin (x)/x for 8.25 radians
0 0 1 1 - Haversine
0 1 0 0 - Square wave
0 1 0 1 - First 10 terms of a square wave
0 1 1 0 - First 4 terms of a square wave
0 1 1 1 - Falling ramp
1 0 0 0 - First 20 terms of a falling ramp
1 0 0 1 - Rising ramp
1 0 1 0 - First 20 terms of a rising ramp
1 0 1 1 - White noise
1 1 0 0 - Modulated white noise
1 1 0 1 - A 3rd, 4th, 5th harmonic chord
1 1 1 0 - 8 cycles of linear rising sine wave
1 1 1 1 - Positive half-cycle sine wave
0 0 - RAM 0
0 1 - RAM 1
1 0 - RAM 2
1 1 - RAM 3
Setting Frequency Register 1: Fixed frequency (CW)
Sweep start frequency
1st FSK frequency
Command Opcode = 1
Bits
7
6
5
4
3
2
1
0
Byte 1 ------->
Most Significant Byte of Frequency
Byte 2 ------->
2nd Significant Byte of Frequency
Byte 3 ------->
3rd Significant Byte of Frequency
Byte 4 ------->
Least Significant Byte of Frequency
Setting Frequency Register 2: Fixed frequency (CW)
Sweep stop frequency
2nd FSK frequency
Command Opcode = 2
Bits
7
6
5
4
3
2
1
0
Byte 1 ------->
Most Significant Byte of Frequency
Byte 2 ------->
2nd Significant Byte of Frequency
Byte 3 ------->
3rd Significant Byte of Frequency
Byte 4 ------->
Least Significant Byte of Frequency
Table C-2. HP E1340A Command Opcodes and Parameter Formats
240 HP E1340A Register-Based Programming
Appendix C
Summary of Contents for E1340A
Page 12: ...Notes 12 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 14: ...14 HP E1340A Arbitrary Function Generator Module User s Manual ...
Page 42: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 42 ...
Page 54: ...Chapter 2 Generating Standard Waveforms with the HP E1340A 54 ...
Page 58: ...Chapter 3 Generating Arbitrary Waveforms with the HP E1340A 58 ...
Page 84: ...Chapter 4 HP E1340A Sweeping and Frequency Shift Keying 84 ...
Page 130: ...Chapter 6 HP E1340A High Speed Operation 130 ...
Page 202: ...202 HP E1340A SCPI Conformance Information Chapter 7 ...
Page 218: ...218 HP E1340A Specifications Appendix A ...
Page 284: ...284 HP E1340A Register Based Programming Appendix C ...
Page 295: ...Index HP E1340A Arbitrary Function Generator User s Manual 295 ...