To make the test connectors
The test connectors connect the logic analyzer to the test equipment.
Table 8
Materials Required
Description
Recommended Part
Qty
BNC (f) Connector
HP 1250-1032
2
Berg Strip, 8-by-2
1
Berg Strip, 1-by-2
1
100
Ω
1% resistor
HP 0698-7212
2
Jumper wire
1
Build the first test connector using a BNC connector and an 8-by-2
section of Berg strip.
a
Solder a jumper wire to all pins on one side of the Berg strip.
b
Solder a jumper wire to all pins on the other side of the Berg strip.
c
c
Solder the center of the BNC connector to the center pin of one row
on the Berg strip.
d
Solder the ground tab of the BNC connector to the center pin of the
other row on the Berg strip.
8-by-2 Connector Setup
Figure 52
Testing, Adjusting, and Troubleshooting the Analyzer
To make the test connectors
156