Example
Consider a situation where you have a squarewave that swings from 0 V to
+60 V; thus, it is too large for the logic inputs
or
the external trigger input.
However, by using a 10:1 divider probe, such as are in the HP 54620A/C, you
can safely apply the squarewave to the external trigger input. You can then
set the threshold for the external trigger to +3.0 V (60 divided by 2 divided
by 10). Then, set an edge trigger for the rising or falling edge as desired.
Signal Margin on External Trigger
The above example uses a 10:1 divider probe because of the large swing of the
analog signal. But, consider the situation where the analog signal swing is 0 to
4 V. The 10:1 divider will reduce this to a 400 mV swing. However, the minimum
input swing required by the analyzer is 500 mV about the threshold. Thus, the
analyzer input will not reliably recognize the signal transitions. The solution is to
use a 1:1 probe. See chapter 7, “Performance Characteristics,” for more
information on the input characteristics of the analyzer.
Using the External Trigger Input as an Edge Trigger
Figure 32
Making Analyzer Measurements
To use the external trigger input
109
Summary of Contents for 54620A
Page 7: ...6 ...
Page 13: ...12 ...
Page 17: ...16 ...
Page 18: ...1 Getting Started ...
Page 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Page 52: ...2 Making Analyzer Measurements ...
Page 121: ...120 ...
Page 122: ...3 Solving Problems ...
Page 127: ...126 ...
Page 128: ...4 Ensuring Accurate Measurements ...
Page 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Page 180: ...6 Replaceable Parts ...
Page 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Page 196: ...7 Performance Characteristics ...
Page 208: ...8 Messages ...
Page 217: ...216 ...
Page 239: ......
Page 241: ...240 ...