To change the time reference position
1
Press
Main/Delayed
.
2
Choose the time reference position from the
Time Reference
softkeys:
•
To set the time reference so that most of the acquired data follows the
trigger, press the
Left
softkey.
The time reference indicator will be one division in from the left side
of the display. The trigger event will be at the same location unless the
delay is set to a value other than 0.
•
To set the time reference so that the acquired data is centered around the
trigger, press the
Center
softkey.
The time reference indicator will be at the middle of the display. The
trigger event will be at the same location unless the delay is set to a
value other than 0.
•
To set the time reference so that most of the acquired data precedes the
trigger, press the
Right
softkey.
The time reference indicator will be one division in from the right side
of the display. The trigger event will be at the same location unless the
delay is set to a value other than 0.
The time reference position sets the initial position of the trigger event within
acquisition memory and on the display, with delay set to 0. The delay setting
sets the specific location of the trigger event with respect to the time
reference position. The time reference setting affects the delayed sweep as
described in the following ways.
The point about which the delayed sweep is expanded changes (as Time/Div
is varied):
•
For a time reference setting of Left, the sweep expands from the
left-hand side of the display.
•
For a time reference setting of Center, the sweep expands about the
center of the display.
•
For a time reference setting of Right, the sweep expands from the
right-hand side of the display.
Making Analyzer Measurements
To change the time reference position
83
Summary of Contents for 54620A
Page 7: ...6 ...
Page 13: ...12 ...
Page 17: ...16 ...
Page 18: ...1 Getting Started ...
Page 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Page 52: ...2 Making Analyzer Measurements ...
Page 121: ...120 ...
Page 122: ...3 Solving Problems ...
Page 127: ...126 ...
Page 128: ...4 Ensuring Accurate Measurements ...
Page 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Page 180: ...6 Replaceable Parts ...
Page 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Page 196: ...7 Performance Characteristics ...
Page 208: ...8 Messages ...
Page 217: ...216 ...
Page 239: ......
Page 241: ...240 ...