![background image](http://html.mh-extra.com/html/hp/53131a/53131a_operating-manual_156805072.webp)
Chapter 1 Getting Started
Making Measurements
1-32
Operating Guide
1
The Counter is now set to make statistics based on 20 measurements.
In this case, the displayed standard deviation value is computed on all
measurements of the 10 MHz signal since no limits were set.
Computing Stats on Filtered Data Only
A special feature of the Counter allows you to use the upper and lower
limits to filter data before statistical processing or computation as shown
in Figure 1-4.
Figure 1-4. Filtering Data Before Statistical Computation
1
Raw Measurements
2
Limits
3
Filtered data (USE: IN LIMIT)
4
Statistics
Run
Freq
Ch 2
Hz
Freq
Ch 2
Hz
Gate
Freq
Ch 2
Gate
1
2
3
4