Chapter 2 Operating Your Universal Counter
Preset Values After Power-Up and *RST
Operating Guide
2-91
2
Preset Values for Functions Accessible Via HP-IB Only
Table 2-8. HP 53131A/132A Preset Values—Accessible Via HP-IB Only
Description
In
Save/Recall
Value at *RST
(HP-IB Reset)
Value at
Power-Up
In
Non-Volatile
Memory
Arming Parameters
Frequency ext stop arm slope
yes
NEG
NEG
no
Totalize ext stop arm slope
yes
NEG
NEG
no
Math, Limit Parameters
Math, limit-test recalculate
yes
OFF
ON
no
Reset limit-test fail count on
INITiate
yes
ON
ON
no
Statistics Parameters
Statistics type
yes
MEAN
MEAN
no
Miscellaneous Measurement
Parameters
Monitor external timebase
no
TRUE
TRUE
no
Interpolator calibration on/off
no
ON
ON
no
Status Reporting Parameters
Event status enable
no
________
0
no
Service request enable
no
________
0
no
OSR enable
no
________
0
no
OSR negative transition filter
no
________
0
no
OSR positive transition filter
no
________
1809
no
QSR enable
no
________
0
no
QSR positive transition filter
no
________
18277
no
QSR negative transition filter
no
________
0
no
Error queue
no
________
cleared
no
Summary of Contents for 53131A
Page 1: ...Operating Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 26: ...xxiv Operating Guide ...
Page 41: ...1 Getting Started ...
Page 77: ...2 Operating Your Universal Counter Operator s Reference ...
Page 171: ...3 Specifications ...