Chapter 3 Specifications
Measurement Specifications (Continued)
Operating Guide
3-11
3
Duty Cycle
Measurement is specified over the full signal range of Channel 1. However, both the positive and negative pulse widths must be greater
than 4 ns.
Results Range:
0 to 1 (e.g. 50% duty cycle would be displayed as .5)
RMS Resolution:
t
res
Totalize
Measurement is specified over the full signal range of Channel 1.
Results Range:
0 to 10
15
Resolution:
± 1 count
Peak Volts
Measurement is specified on Channels 1 and 2 for DC signals; or for AC signals of frequencies between 100 Hz and 30 MHz with peak-to-
peak amplitude greater than 100 mV. (The measurement will continue to operate up to 225 MHz, though results are for indication only.)
Results Range:
–5.1 V to +5.1 V
Resolution:
10 mV
Systematic Uncertainty for AC signals:
25 mV + 10% of V
for DC signals:
25 mV + 2% of V
Use of the input attenuator multiplies all voltage specifications (input range, results range, resolution and systematic uncertainty) by a
nominal factor of 10. For example with AC signals, the Systematic Uncertainty becomes: 250 mV + 10% of V.
Measurement Specifications (Continued)
t
res
2
2
Trigger Error
2
×
(
)
+
(
)
1
Duty Cycle
2
+
(
)
×
Frequency
×
53131A 53132A
750 ps 300 ps
--------------------------------------------------------------------------------------------
Summary of Contents for 53131A
Page 1: ...Operating Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 26: ...xxiv Operating Guide ...
Page 41: ...1 Getting Started ...
Page 77: ...2 Operating Your Universal Counter Operator s Reference ...
Page 171: ...3 Specifications ...
Page 196: ... Manual Part Number 53131 90055 Printed in U S A MAY 1999 ...