Chapter 2 Operating Your Universal Counter
Using the Gate & External Arm Menu Key
2-24
Operating Guide
2
To Use the Gate and External Arm
Example Procedure for Gate and External Arm
For demonstration purposes, you want to set up the Counter so that it
uses an external arm to start a measurement. You want to make a
Totalize measurement on the positive edges (that is, count the number of
positive edges) of a pulse-train signal applied to channel 1. Also, you want
to perform the measurement during an interval of 5 milliseconds as shown
in the following figure.
Perform the following procedure to accomplish this task.
1
Turn on the Counter, and press Stop/Single to put the Counter into the
single measurement mode.
2
Connect the pulse-train signal to channel 1 input, and press Other Meas
key until
TOTALIZE 1
is displayed.
3
Connect an external TTL signal to the Counter’s rear-panel Ext Arm
connector.
4
Press Gate & ExtArm key.
GATE: TIME
is displayed.
5
Press any one of the arrow keys until
GATE: EXTERNL
is displayed.
6
Press Gate & ExtArm key.
START: POS
is displayed. (If
START: NEG
is displayed instead, then
press any one of the arrow keys to toggle to the next state—
POS
).
Gate
Ext
Arm
CHANNEL 1
5 ms
Ext
Arm
Summary of Contents for 53131A
Page 1: ...Operating Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 26: ...xxiv Operating Guide ...
Page 41: ...1 Getting Started ...
Page 77: ...2 Operating Your Universal Counter Operator s Reference ...
Page 171: ...3 Specifications ...
Page 196: ... Manual Part Number 53131 90055 Printed in U S A MAY 1999 ...