Chapter 2 Operating Your Universal Counter
Preset Values After Power-Up and *RST
2-86
Operating Guide
2
Phase Arming Parameters
Phase arm
yes
IMMediate
AUTO
no
Phase ext start arm slope
yes
POS
POS
no
Pulse, Rise, Fall, Dutycycle
Arming Parameters
Pulse Arming Parameters
Pulse start arm
yes
IMMediate
AUTO
no
Pulse ext start arm slope
yes
POS
POS
no
Time Interval Arming
Parameters
Time Interval start arm
yes
IMMediate
AUTO
no
Time Interval ext start arm slope
yes
POS
POS
no
Time Interval start arm delay
yes
IMMediate
NONE
no
Time Interval start arm delay
time
yes
100 E-9s
100 E-9s
no
Time Interval start arm delay
events
yes
1
1
no
Time Interval stop arm
yes
IMMediate
AUTO
no
Time Interval ext stop arm slope
yes
POS
POS
no
Time Interval stop arm delay
yes
IMMediate
NONE
no
Time Interval stop arm delay
time
yes
100 E-9s
100 E-9s
no
Time Interval stop arm delay
events
yes
1
1
no
Input Coupling Parameters
Channel 1 input coupling
yes
AC
AC
no
Channel 2 input coupling
yes
AC
AC
no
Table 2-7B. HP 53132A (S/N 3646 and above) Preset Values (continued)
Description
In
Save/Recall
Value at *RST
(HP-IB Reset)
Value at
Power-Up
In
Non-Volatile
Memory
Summary of Contents for 53131A
Page 1: ...Operating Guide HP 53131A 132A 225 MHz Universal Counter ...
Page 2: ......
Page 26: ...xxiv Operating Guide ...
Page 41: ...1 Getting Started ...
Page 77: ...2 Operating Your Universal Counter Operator s Reference ...
Page 171: ...3 Specifications ...