Chapter 2 Operating Your Universal Counter
Using Entry/Select (Arrow) Keys
Operating Guide
2-9
2
To Use During State Changing (ON/OFF, LO/MED/HI,
etc.)
•
Press any of the arrow keys to toggle or change to the next state of the
parameters found in the following menus:
–
Gate & ExtArm
–
Limit Modes
–
Scale & Offset
–
Stats
–
Trigger/Sensitivity
–
50
Ω
/1M
Ω
–
DC/AC
–
X10 Attenuate
–
100kHz Filter
–
Save & Print
–
Utility Menu
•
The Enter and +/
−
keys are ignored during the state changing
operating mode.
To Use on Prompted Event Messages (SET OFFSET ?,
CAL: OFFS n ?, TEST: ALL?, etc.)
•
For
SET OFFSET ?
, and all the
TEST: ?
prompted event messages,
press Enter key to start the event.
•
For
CAL: OFFS n ?
and
GAIN n ?
, press Enter key to display
instructions. Press Enter key again to start the event.
To Use on Prompted Help Messages (MATH HELP ?,
PRINT HELP ?)
•
When
MATH HELP ?
is displayed, press Enter key to display the
“
(MEAS x SCALE) + OFFS = RESULT
” message.
•
When
PRINT HELP ?
is displayed, press Enter key to display the
“
CONFIGURE PORT ON UTILITY MENU; REMEMBER TO SAVE
SETUP FIRST
” instruction.