To replace the Reference Clock cable
1
Remove the module from the mainframe. Follow the procedure "To remove the
module" on page 6-2.
2
Unplug the faulty Reference Clock cable from the circuit board REF CLK OUT (J12)
or REF CLK IN (J13) connector.
3
Using a 1/4 inch hollow shaft nutdriver, remove the nut that holds the cable to the
module panel insert.
4
Remove and replace the faulty Reference Clock cable. Insert the connector nut and
gently tighten the nut using the nutdriver.
5
Plug the replacement cable into the board REF CLK OUT (J12) or REF CLK IN (J13).
connector.
6
Follow the procedure "To replace the module" to reinstall the module into the
mainframe.
Replacing Assemblies
To replace the Reference Clock cable
6–6