To test the time interval accuracy
Testing the time interval accuracy does not check a specification, but does check the
following:
•
125 MHz oscillator
Multicard modules must be reconfigured as one-card modules for this test.
This test verifies that the 125 MHz timing acquisition synchronizing oscillator is
operating within limits.
Equipment Required
Equipment
Critical Specifications
Recommended
Model/Part
Pulse Generator
100 MHz 3.5 ns pulse width, <600 ps rise time
HP 8131A Option 020
Function Generator
Accuracy
≤
(5)(10
−
6
)
×
frequency
HP 3325B Option 002
SMA Cable
HP 8120-4948
Adapter
BNC(m)-SMA(f)
HP 1250-2015
BNC Test Connector,
6x2
Set up the equipment
1
Turn on the equipment required and the logic analyzer. Let them warm up for
30 minutes if you have not already done so.
2
Set up the pulse generator according to the following table.
Pulse Generator Setup
Channel 1
Period
Mode
EXT TRIG
Delay: 0 ps
5
µ
s
TRIG
Slope: Positive
Width: 2.5
µ
s
THRE: 1.0 V
High:
−
0.9 V
Low:
−
1.7 V
COMP: Disabled
(LED off)
3–51
Summary of Contents for 16555A
Page 4: ...The HP 16555A D Logic Analyzer iii ...
Page 15: ...1 8 ...
Page 97: ...3 70 ...
Page 98: ...4 Calibrating ...
Page 102: ...Troubleshooting Flowchart 1 Troubleshooting To use the flowcharts 5 3 ...
Page 103: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 5 4 ...
Page 104: ...Troubleshooting Flowchart 3 Troubleshooting To use the flowcharts 5 5 ...
Page 105: ...Troubleshooting Flowchart 4 3 Troubleshooting To use the flowcharts 5 6 ...
Page 117: ...5 18 ...
Page 125: ...6 8 ...
Page 131: ...7 6 ...
Page 132: ...8 Block Level Theory 8 2 Self Tests Description 8 6 Theory of Operation ...