To run the self-tests
Self-tests for the module identify the correct operation of major functional areas of the
module. The self-tests consist of Board Verification tests and Chip Verification tests. You can
run all self-tests without accessing the interior of the instrument. If a self-test fails, the
troubleshooting flowcharts instruct you to change a card or cable of the module.
1
Disconnect all inputs, then turn on the power switch.
2
In the System Configuration menu, touch Configuration. In the pop-up menu, touch
Test.
3
In the Test menu, touch the box labeled Touch box to Load Test System.
4
In the test system screen, touch Test System. Select the 1M Sample LA (HP 16555A)
or 2M Sample LA (HP 16555D) module to be tested.
Troubleshooting
To run the self-tests
5–7