3
Set up the oscilloscope.
a
Select Setup, then select Default Setup.
b
Configure the oscilloscope according to the following table.
Oscilloscope Setup
Acquisition
Display
Trigger
[Shift]
∆
Time
Averaging: On
# of averages: 16
Graticule
Graphs: 2
Level: -250 mV
Stop src: channel 2 [Enter]
Channel 1
Channel 2
Define meas
Alternate Scale
Attenuation: 20.00:1
Scale: 200 mV/div
Offset:
−
1.300 V
Alternate Scale
Attenuation: 20.00:1
Scale: 200 mV/div
Offset: -1.300 V
Thresholds: user-defined
Units: Volts
Upper: -980 mV
Middle: -1.30 V
Lower: -1.62 V
Set up the logic analyzer
1
Set up the Configuration menu.
a
In the System Configuration menu, touch System, then select 1M Sample LA (2M
Sample LA for HP 16555D).
b
In the Configuration menu, assign all pods to Machine 1. To assign the pods, touch the
pod fields, then select Machine 1.
c
In the Analyzer 1 box, touch the Type field, then select State.
Testing Performance
To test the single-clock, single-edge, state acquisition
3–19
Summary of Contents for 16555A
Page 4: ...The HP 16555A D Logic Analyzer iii ...
Page 15: ...1 8 ...
Page 97: ...3 70 ...
Page 98: ...4 Calibrating ...
Page 102: ...Troubleshooting Flowchart 1 Troubleshooting To use the flowcharts 5 3 ...
Page 103: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 5 4 ...
Page 104: ...Troubleshooting Flowchart 3 Troubleshooting To use the flowcharts 5 5 ...
Page 105: ...Troubleshooting Flowchart 4 3 Troubleshooting To use the flowcharts 5 6 ...
Page 117: ...5 18 ...
Page 125: ...6 8 ...
Page 131: ...7 6 ...
Page 132: ...8 Block Level Theory 8 2 Self Tests Description 8 6 Theory of Operation ...