![horiba LO-300 Instruction Manual Download Page 72](http://html1.mh-extra.com/html/horiba/lo-300/lo-300_instruction-manual_2162291072.webp)
Troubleshooting
58
Message list
Table 10 Message list
No.
Message
Explanation
Countermeasure
Refer
to
page
1
Factory preset data is not
available. PLEASE
CONTACT TO
YOUR LOCAL AGENT.
Displayed when any abnormality is
found at power on. Either because the
data required for monitoring is absent
or broken, normal monitoring cannot
be performed.
Unexpected abnormality may
occur. Contact us.
2
Factory preset data is not
available. Current data is
used. PLEASE CONTACT TO
YOUR
LOCAL AGENT.
Displayed when any abnormality
occurs when the operation of
“returning to factory default”. Since the
factory default data required for
monitoring has been broken, the
current setting data is used.
Unexpected abnormality may
occur. Contact us.
3
NO LOGGING DATA.
No history is available.
4
REPLACE LASER MODULE Laser needs to be replaced.
When this message appears,
contact us.
5
ERROR: Must be ALARM
=
WARNING.
The condition that “Detection area
(ALARM)”
Detection area
(WARNING)” is not satisfied.
Correctly set to satisfy “Detection
area (ALARM)”
Detection area
(WARNING)”.
6
ERROR: It is invalid date.
Unreal date has been set.
Displayed when an unreal date is
registered in the “Date setting”.
7
ERROR: Data saving is
failed.
Data saving has been failed.
Displayed when the data
registration is failed for data
setting. When the error persists
even after the reset operation,
contact us.
-
8
ERR: Screen = Title color
“Background color of the entire
screen” and “Title font color” is the
same.
Set a different color.
9
ERR: Screen = Warning color
“Background color of the entire
screen” and “Oil film warning color” is
the same.
Set a different color.
10 ERR: Screen = Alarm color
“Background color of the entire
screen” and “oil film alarm color” is the
same.
Set a different color.
11 ERR: Screen = Bar color
“Background color of the entire
screen” and the “bar color of oil film
monitoring” is the same.
Set a different color.
12 ERR: Button = Font color
“Background color of button” and
“button font color” is the same.
Set a different color.
13
ERROR: Must be
UPPER
LOWER
The condition “upper limit level”
“lower limit level is not satisfied.
Set upper limit level
lower limit
level
14
Oil level is not assigned
between the upper limit and
the lower limit. Is it O.K. to
start oil detection?
The condition “lower limit level <
detection level
upper limit level” is
not satisfied. Do you want to start the
monitoring in this condition?
Touching “YES” can start
monitoring, but correct judgment
cannot always be performed.
15
WARNING: Must be OIL
LEVEL
UPPER
The condition “detection level
upper
limit level” is not satisfied.
Correctly set so that “detection
level
upper limit level” is
satisfied.