Overview
9
Factory default setting
Table 5 shows the setting at the factory shipment.
The following setting can be changed by operating the touch panel on the main body display.
Table 5 Factory default setting
Item
Default value
Description
Detection level setting
13 mA
When this value is exceeded, oil film is identified.
Detection area setting (Warning)
0
The area detection function is invalid.
Detection area setting (Alarm)
0
The area detection function is invalid.
Continuous judgment time setting
(Warning)
10 s
-
Continuous judgment time setting
(Alarm)
10 s
-
Contact output setting
Alarm
The output of “ALARM” or “WARNING” can be selected.
Contact output hold time setting
0 s
Alarm output is released soon after the normal status is
restored.
Lower limit error judgment setting
OFF
-
Upper limit setting
22 mA
Any value exceeding this value will be judged as upper
limit error level.
Lower limit setting
5 mA
Any value falling below this value will be judged as lower
limit error level (If Lower error judge setting is ON).
Multi-Trans function setting
OFF
-
Laser replacement
Contact output time setting
0 hour 0 minute
-